Tip# 115: Adduct formation in LC-MS (esp. ESI)

Almost everything you analyze by Electrospray ionization mass spectrometry will create an adduct with something in the system. Normally, hydrogen is the most common adduct formed (M+1), but other chemicals, often in trace amounts may form adducts with your sample too. Sometimes we can take advantage of this fact and introduce our own adduct into the system (post column) to increase signal sensitivity or help us isolate one signal from another (the addition of an adduct can sometimes increase the signal seen for one species, but not the other).

One of my favorite elements to form an adduct with is sodium (Na+). Two common forms are sodium citrate and sodium acetate. Both have PKA’s between 3 and 6 so a variety of buffered solutions can be prepared for use. However, in LC-MS it is very important that we keep the concentration of sodium as low as possible so as to not clog the mass detector or suppress ionization completely (and see nothing BUT Sodium for weeks …). My suggestion is to prepare the buffers such that the solution is below 3 mM in concentration. The lowest concentration should be used that yields reproducible results. Only use the highest purity buffers (some manufacturer’s use names such as “ultra” to describe them) when preparing these ‘doping’ solutions for post-column addition and be sure and filter them through a 0.2 micron filter before use. A syringe pump can be used to deliver the solution during the run. A low flow rate should be used to infuse the adduct solution into the main inlet of the detector. Make sure you have a simple way of controlling the pump through the system (e.g. ‘On’ / ‘Off’, contact closure) so the flow can be turned off when you are not acquiring data.

Ammonium (NH4) is another popular adduct to add to the system, Often in the form of ammonium acetate. Whichever adduct you use in your system, always start off testing as low a concentration as possible. Monitor the baseline carefully for noise and also to see if the addition of the compound is suppressing or enhancing the signal generated for your compound. Careful use of adducts in your system can provide you with another means to selectively enhance the signal of some compounds without changing the original chromatography method.

> Bill Letter, 11/17/2011


HINT's & TIPS Back to the Table of Contents

   WB01343_.gif (1341 bytes)

Chiralizer™ Services, L.L.C., Newtown, PA   USA

Copyright 2011, Chiralizer™ Services, L.L.C. All Rights Reserved. Chiralizer, Column Swapper & LC Spiderling are trademarks of Chiralizer Services, Newtown, PA   USA. *These "Hints & Tips" are protected by copyright and are not to be copied or duplicated in whole or part without the written permission of Chiralizer Services, LLC.